Table 1 Summary of the method validation results

From: Rapid, simple, and clinically applicable high-performance liquid chromatography method for clinical determination of plasma colistin concentrations

Parameter

Result

Concentration range (μg/mL)

0.10–8.0

Regression equationa

y = 0.6496× − 0.0141

SE of slopeb

0.0230

SE of interceptb

0.0019

Determination coefficient (r2)

0.9999

Intra-day precision (%)c

2.69

Inter-day precision (%)c

4.01

Intra-day accuracy (%)c

103.9

Inter-day accuracy (%)c

105.2

Lower limit of quantification (μg/mL)

0.10

Limit of detection (μg/mL)

0.025

Stability

acceptable

Robustness

acceptable

  1. ay = ax + b, where y is the ratio of the summed peak areas of colistin A and B to that of the internal standard, and x is colistin concentration (μg/mL)
  2. bSE, standard error (n = 3)
  3. cMean of four concentrations (0.10, 0.50, 2.0, and 4.0 μg/mL)