Skip to main content

Table 2 Intraday and interday assay precision and accuracy for fluconazole in serum samples

From: A sensitive method for analyzing fluconazole in extremely small volumes of neonatal serum

Conditions

Concentration (μg/mL)

Accuracy (%)

Precision (%)

Intraday

0.01

−4.6

3.1

0.1

4.7

4.6

1.0

−0.4

2.1

10.0

7.4

3.4

50.0

5.0

6.0

100.0

−7.7

5.8

Interday

0.01

−4.4

4.3

0.1

0.8

2.7

1.0

3.0

0.3

10.0

3.2

2.5

50.0

−2.5

2.5

100.0

−0.3

5.4