Table 2 UPLC-MS/MS conditions and settings

From: Evaluation of the permeation of antineoplastic agents through medical gloves of varying materials and thickness and with varying surface treatments

Analyte

Monitor ion [m/z]

Columna

Flow [mL/min]

Composition (solvent A/solvent B)b

LOQ [ng/mL]

ETPc

589.5 > 229.1

A

0.3

30/70 (isocratic)

50

CPAd

261.1 > 139.7

A

0.3

30/70 (isocratic)

40

DXR

544.4 > 397.2

A

0.3

0–3

3–4.5

4.5–5.5

min

min

min

90/10

90/10 → 10–90

90/10

50

PTX

854.6 > 286.0

A

0.3

0–3

3–4.5

4.5–5.5

min

min

min

70/30

70/30 → 10–90

70/30

200

5FU

128.6 > 85.6e

B

0.3

50/50 (isocratic)g

100

Gemcitabinef

264.1 > 111.6

-

  1. aColumn A, ACQUITY UPLC BEH® Shield RP18 (1.7 μm, 2.1 × 100 mm); column B, ACQUITY UPLC BEH® AMIDE (1.7 μm, 2.1 × 150 mm)
  2. bSolvent A, 0.1% (v/v) formic acid in water; solvent B, 0.1% (v/v) formic acid in acetonitrile
  3. cAlso used as internal standard for CPA, DXR, and PTX
  4. dAlso used as internal standard for ETP
  5. e5FU was detected by negative ion mode whereas the other compounds were detected by positive ion mode
  6. fInternal standard for 5FU
  7. gWater/acetonitrile = 50/50 (without 0.1% formic acid)