Skip to main content

Table 2 Intra- and inter-day precisions of the assay for determining colistin concentration in the control serum

From: Rapid, simple, and clinically applicable high-performance liquid chromatography method for clinical determination of plasma colistin concentrations

Colistin concentration (μg/mL)

Intra-day

Inter-day

Precision (%)

Accuracy (%)

Precision (%)

Accuracy (%)

High QC (4.00)

0.87

102.0

1.97

102.3

Middle QC (2.00)

2.64

99.0

4.19

101.9

Low QC (0.50)

3.74

100.7

3.71

101.4

LLOQ (0.10)

3.50

113.8

6.17

115.0

  1. QC, quality control; LLOQ, lower limit of quantitation